Known Issues

  • High resolution diffractometer (HRD)
    • blank images at the end of datasets with small oscillation range (0.01~) and big number of images (5000~)
    • detector cover stucks sometime
    • long exposure time (more than 3-6 sec) might cause saturated images
      • should be solved but might happen
  • CATS
    • mounting is not stable (1-2 pins/dewar)
    • the gripper sometimes stops working just above the lid after putting the vial in the dewar
      • open CATS debug panel
      • repeat power off/on till the robot is on
      • click move to home position
      • in case close the current GUI and open a new one
  • micro diffractometer MkII (MD2)
    • 1-click alignment/vertical sample translation is disabled
      • go to options->software setup->interval angle for 3-click
      • set 0 for the above parameter
      • click the sample 3 time at the same point to simulate 1-click alignment
    • 5 and 10 micron aperture is exactly 2.400 and 1.200 mm above the 20 micron aperture
-- Main.TakashiTomizaki - 20 Jul 2011