JUNGFRAU: adJUstiNg Gain detector FoR the Aramis User station

JUNGFRAU is a state-of-the-art, two-dimensional pixel detector designed for high-performance photon science applications at Free Electron Lasers (FEL) and synchrotron light sources. It was primarily developed for the SwissFEL machine at the Paul Scherrer Institute in Switzerland. JUNGFRAU is tailored to meet the stringent demands of these advanced facilities, offering exceptional performance in terms of dynamic range, noise reduction, and readout speed, making it ideal for cutting-edge experiments.

JUNGFRAU 1.0 readout ASIC corner, without sensor.

Design and performance characteristics

  • Two-dimensional, modular/tilable, vacuum-compatible imaging detector
  • 75 µm pixel size, comparable to photon counting detectors
  • Single photon sensitivity
  • High dynamic range with up to 10,000 12 keV photons/pulse/pixel
  • Low noise over entire dynamic range
  • High detection linearity over the full dynamic (non-linearity < 1%)
  • Image rate > 2.2kHz continuous (up to 250 keV in burst mode)
  • Energy range > 2 keV @ single photon sensitivity (with standard silicon sensors)


Implementation
JUNGFRAU is a charge-integrating detector designed to support applications involving high photon rates, making it particularly well-suited for high-intensity photon sources like X-ray Free Electron Lasers (XFEL) and synchrotron facilities. A key feature of JUNGFRAU is its three-stage gain-switching preamplifier embedded within each pixel, which dynamically adjusts the gain based on the amount of charge deposited. This mechanism, similar to that used in other advanced detectors like AGIPD and GOTTHARD, ensures single-photon sensitivity while maintaining a dynamic range spanning four orders of magnitude. JUNGFRAU has demonstrated single-photon detection down to energies as low as 1.5 keV under short, XFEL-compatible exposure times.

Geometry
The JUNGFRAU chip consists of 256 × 256 pixels, each measuring 75 µm × 75 µm. These chips are assembled into arrays of 2 × 4 to form modules measuring 4 cm × 8 cm, providing a total of approximately 500 kPixels per module. The chips are typically paired with 320 µm thick silicon sensors, although 450 µm sensors are available upon request, largely following the design principles of the EIGER detector. Multi-module systems, with up to 16 Mpixels (comprising 32 modules), are available for various experimental setups at SwissFEL’s end stations and at several synchrotron light sources.

Readout
The JUNGFRAU readout system is modeled after the GOTTHARD strip detector system, optimized for XFEL applications. It is fully modular, with individual boards requiring only a 12 V power supply, a copper Ethernet connection for controls, and a 10 Gbit fiber link for data transmission. The system is vacuum-compatible, using a slotted aluminum flange that is glued onto the flat part of the readout board for the vacuum interface. JUNGFRAU supports operation at frame rates ranging from a few Hz to 2.2 kHz with dead time-free performance, easily meeting readout demands of SwissFEL. Additionally, it enables high-rate synchrotron experiments with an exceptionally high linear count rate capability of over 25 MHz per pixel (at 12 keV photon energy with 2.2 kHz readout).

Components of a JUNGFRAU module. From left to right, in the first row: Sensor, ASICs, High Density Interconnect, aluminum holder, readout board. In the background you can see the sensor wafer and the ASIC wafers.

Cameras
Using single modules as building blocks, several detector cameras are available with customizable configurations. Single module (500 kPixels), 1 Mpixel, and 1.5 Mpixel systems with vertically stacked modules, as well as multiple 4 Mpixel and 9 Mpixel cameras have been produced and are commonly used at several synchrotron and FEL facilities. A 4.5 Mpixel linear module array for X-ray emission spectroscopy at the ALVRA SwissFEL endstation has been commissioned and successfully operated during the SwissFEL pilot experiment phase in December 2017 (link to ALVRA pilot, link to BERNINA pilot). Two 16 Mpixel cameras for the BERNINA and ALVRA endstations were installed and commissioned in 2018.

 


  • X-ray spectroscopy: X-ray absorption (XAS), X-ray emission (XES), resonant inelastic X-ray scattering (RIXS), X-ray diffuse scattering
  • Nanocrystallography
  • X-ray diffraction
  • X-ray pump-probe <diffraction and scattering: time-resolved X-ray diffraction (trXRD), time-resolved resonant X-ray diffraction (trRXRD) and time-resolved resonant inelastic X-ray scattering (trRIXS)
  • Protein crystallography, specifically high-rate applications
  • Hinger V, Barten R, Baruffaldi F, Bergamaschi A, Borghi G, Boscardin M, et al.
    Resolving soft X-ray photons with a high-rate hybrid pixel detector
    Frontiers in Physics. 2024; 12: 1352134 (10 pp.). https://doi.org/10.3389/fphy.2024.1352134
    DORA PSI
  • Leonarski F, Brückner M, Lopez-Cuenca C, Mozzanica A, Stadler H-C, Matěj Z, et al.
    Jungfraujoch: hardware-accelerated data-acquisition system for kilohertz pixel-array X-ray detectors
    Journal of Synchrotron Radiation. 2023; 30: 227-234. https://doi.org/10.1107/S1600577522010268
    DORA PSI
  • Leonarski F, Nan J, Matej Z, Bertrand Q, Furrer A, Gorgisyan I, et al.
    Kilohertz serial crystallography with the JUNGFRAU detector at a fourth-generation synchrotron source
    IUCrJ. 2023; 10(6): 729-737. https://doi.org/10.1107/S2052252523008618
    DORA PSI
  • Schreiner S, Rauch C, Akstaller B, Bleuel P, Fröjdh E, Martynenko AS, et al.
    Electromagnetic pulse protective shielding for digital x-ray detectors
    Review of Scientific Instruments. 2023; 94(7): 075106 (8 pp.). https://doi.org/10.1063/5.0160120
    DORA PSI
  • Sikorski M, Ramilli M, de Wijn R, Hinger V, Mozzanica A, Schmitt B, et al.
    First operation of the JUNGFRAU detector in 16-memory cell mode at European XFEL
    Frontiers in Physics. 2023; 11: 1303247 (14 pp.). https://doi.org/10.3389/fphy.2023.1303247
    DORA PSI
  • Fröjdh E, Baruffaldi F, Bergamaschi A, Carulla M, Dinapoli R, Greiffenberg D, et al.
    Detection of MeV electrons using a charge integrating hybrid pixel detector
    Journal of Instrumentation. 2022; 17(12): C12004 (8 pp.). https://doi.org/10.1088/1748-0221/17/12/C12004
    DORA PSI
  • Fröjdh E, Abrahams JP, Andrä M, Barten R, Bergamaschi A, Brückner M, et al.
    Electron detection with CdTe and GaAs sensors using the charge integrating hybrid pixel detector JUNGFRAU
    Journal of Instrumentation. 2022; 17: C01020 (12 pp.). https://doi.org/10.1088/1748-0221/17/01/C01020
    DORA PSI
  • Hinger V, al Haddad A, Barten R, Bergamaschi A, Brückner M, Carulla M, et al.
    Advancing the JUNGFRAU detector toward low-energy X-ray applications
    Journal of Instrumentation. 2022; 17(9): C09027 (7 pp.). https://doi.org/10.1088/1748-0221/17/09/C09027
    DORA PSI
  • Meyer M, Andrä M, Barten R, Bergamaschi A, Brückner M, Busca P, et al.
    Observation of radiation damage in CdTe Schottky sensors created by 20 keV photons
    Journal of Instrumentation. 2022; 17(6): P06035 (22 pp.). https://doi.org/10.1088/1748-0221/17/06/P06035
    DORA PSI
  • Greiffenberg D, Andrä M, Barten R, Bergamaschi A, Brückner M, Busca P, et al.
    Characterization of chromium compensated GaAs sensors with the charge-integrating JUNGFRAU readout chip by means of a highly collimated pencil beam
    Sensors. 2021; 21(4): 1550 (22 pp.). https://doi.org/10.3390/s21041550
    DORA PSI
  • Mariette C, Lorenc M, Cailleau H, Collet E, Guérin L, Volte A, et al.
    Strain wave pathway to semiconductor-to-metal transition revealed by time-resolved X-ray powder diffraction
    Nature Communications. 2021; 12: 1239 (11 pp.). https://doi.org/10.1038/s41467-021-21316-y
    DORA PSI
  • Fröjdh E, Wennmacher JTC, Rzepka P, Mozzanica A, Redford S, Schmitt B, et al.
    Discrimination of aluminum from silicon by electron crystallography with the Jungfrau detector
    Crystals. 2020; 10(12): 1-14. https://doi.org/10.3390/cryst10121148
    DORA PSI
  • Fröjdh E, George SP, Kodaira S, Lopez-Cuenca C, Mozzanica A, Pinsky LS, et al.
    Measurement of charge deposition from heavy ions with the charge integrating JUNGFRAU detector
    Journal of Instrumentation. 2020; 15(5): C05044 (6 pp.). https://doi.org/10.1088/1748-0221/15/05/C05044
    DORA PSI
  • Leonarski F, Mozzanica A, Brückner M, Lopez-Cuenca C, Redford S, Sala L, et al.
    JUNGFRAU detector for brighter x-ray sources: solutions for IT and data science challenges in macromolecular crystallography
    Structural Dynamics. 2020; 7(1): 014305 (13 pp.). https://doi.org/10.1063/1.5143480
    DORA PSI
  • Martiel I, Mozzanica A, Opara NL, Panepucci E, Leonarski F, Redford S, et al.
    X-ray fluorescence detection for serial macromolecular crystallography using a JUNGFRAU pixel detector
    Journal of Synchrotron Radiation. 2020; 27: 329-339. https://doi.org/10.1107/S1600577519016758
    DORA PSI
  • Redford S, Andrä M, Barten R, Bergamaschi A, Brückner M, Chiriotti S, et al.
    First full dynamic range scan of the JUNGFRAU detector performed at an XFEL with an accurate intensity reference
    Journal of Instrumentation. 2020; 15(2): C02025 (8 pp.). https://doi.org/10.1088/1748-0221/15/02/C02025
    DORA PSI
  • Greiffenberg D, Andrä M, Barten R, Bergamaschi A, Busca P, Brückner M, et al.
    Characterization of GaAs:Cr sensors using the charge-integrating JUNGFRAU readout chip
    Journal of Instrumentation. 2019; 14(05): P05020 (17 pp.). https://doi.org/10.1088/1748-0221/14/05/P05020
    DORA PSI
  • Tolstikova A, Levantino M, Yefanov O, Hennicke V, Fischer P, Meyer J, et al.
    1 kHz fixed-target serial crystallography using a multilayer monochromator and an integrating pixel detector
    IUCrJ. 2019; 6(5): 927-937. https://doi.org/10.1107/S205225251900914X
    DORA PSI
  • Leonarski F, Redford S, Mozzanica A, Lopez-Cuenca C, Panepucci E, Nass K, et al.
    Fast and accurate data collection for macromolecular crystallography using the JUNGFRAU detector
    Nature Methods. 2018; 15(10): 799-804. https://doi.org/10.1038/s41592-018-0143-7
    DORA PSI
  • Mozzanica A, Andrä M, Barten R, Bergamaschi A, Chiriotti S, Brückner M, et al.
    The JUNGFRAU detector for applications at synchrotron light sources and XFELs
    Synchrotron Radiation News. 2018; 31(6): 16-20. https://doi.org/10.1080/08940886.2018.1528429
    DORA PSI
  • Redford S, Andrä M, Barten R, Bergamaschi A, Brückner M, Dinapoli R, et al.
    First full dynamic range calibration of the JUNGFRAU photon detector
    Journal of Instrumentation. 2018; 13(1): C01027 (8 pp.). https://doi.org/10.1088/1748-0221/13/01/C01027
    DORA PSI
  • Redford S, Andrä M, Barten R, Bergamaschi A, Brückner M, Chiriotti S, et al.
    Operation and performance of the JUNGFRAU photon detector during first FEL and synchrotron experiments
    Journal of Instrumentation. 2018; 13(11): C11006 (9 pp.). https://doi.org/10.1088/1748-0221/13/11/C11006
    DORA PSI
  • Milne CJ, Schietinger T, Aiba M, Alarcon A, Alex J, Anghel A, et al.
    SwissFEL: the Swiss X-ray free electron laser
    Applied Sciences. 2017; 7(7): 720 (57 pp.). https://doi.org/10.3390/app7070720
    DORA PSI
  • Rajendran R, Mochi I, Helfenstein P, Mohacsi I, Redford S, Mozzanica A, et al.
    Towards a stand-alone high-throughput EUV actinic photomask inspection tool: RESCAN
    In: Sanchez MI, Ukraintsev VA, eds. Metrology, inspection, and process control for microlithography XXXI. Vol. 10145. Proceedings of SPIE. Bellingham, Washington: SPIE; 2017:101450N. https://doi.org/10.1117/12.2258379
    DORA PSI
  • Jungmann-Smith JH, Bergamaschi A, Brückner M, Cartier S, Dinapoli R, Greiffenberg D, et al.
    Towards hybrid pixel detectors for energy-dispersive or soft X-ray photon science
    Journal of Synchrotron Radiation. 2016; 23(2): 385-394. https://doi.org/10.1107/S1600577515023541
    DORA PSI
  • Mozzanica A, Bergamaschi A, Brueckner M, Cartier S, Dinapoli R, Greiffenberg D, et al.
    Characterization results of the JUNGFRAU full scale readout ASIC
    Journal of Instrumentation. 2016; 11(2): C02047 (8 pp.). https://doi.org/10.1088/1748-0221/11/02/C02047
    DORA PSI
  • Redford S, Bergamaschi A, Brückner M, Cartier S, Dinapoli R, Ekinci Y, et al.
    Calibration status and plans for the charge integrating JUNGFRAU pixel detector for SwissFEL
    Journal of Instrumentation. 2016; 11(11): C11013 (8 pp.). https://doi.org/10.1088/1748-0221/11/11/C11013
    DORA PSI
  • Bergamaschi A, Cartier S, Dinapoli R, Greiffenberg D, Jungmann-Smith JH, Mezza D, et al.
    Looking at single photons using hybrid detectors
    Journal of Instrumentation. 2015; 10(1): C01033 (13 pp.). https://doi.org/10.1088/1748-0221/10/01/C01033
    DORA PSI
  • Jungmann-Smith JH, Bergamaschi A, Brückner M, Cartier S, Dinapoli R, Greiffenberg D, et al.
    Radiation hardness assessment of the charge-integrating hybrid pixel detector JUNGFRAU 1.0 for photon science
    Review of Scientific Instruments. 2015; 86(12): 123110 (9 pp.). https://doi.org/10.1063/1.4938166
    DORA PSI
  • Jungmann-Smith JH, Bergamaschi A, Cartier S, Dinapoli R, Greiffenberg D, Johnson I, et al.
    JUNGFRAU 0.2: prototype characterization of a gain-switching, high dynamic range imaging system for photon science at SwissFEL and synchrotrons
    Journal of Instrumentation. 2014; 9(12): P12013 (12 pp.). https://doi.org/10.1088/1748-0221/9/12/P12013
    DORA PSI
  • Mozzanica A, Bergamaschi A, Cartier S, Dinapoli R, Greiffenberg D, Johnson I, et al.
    Prototype characterization of the JUNGFRAU pixel detector for SwissFEL
    Journal of Instrumentation. 2014; 9(5): C05010 (9 pp.). https://doi.org/10.1088/1748-0221/9/05/C05010
    DORA PSI