Parameters
Sample environment- See special page for information about the Sample Environment
- Guide hall, at the cold neutron guide RNR12, supermirrorguide (m=2, height 12 cm, width 2 cm)
| Monochromators | |
|---|---|
| Pyrolytic Graphite (002) (standard) |
vertically focusing 5 crystals of 2.5 cm height |
| Germanium (311, 511, 400) (optional) |
vertically focusing 9 crystals of 1.4 cm height |
- Any wavelength between 2.3 Å and 5 Å is possible, standard wavelengths (PG002 monochromator) are 2.45 Å, 3.8 Å, 4.2 Å and 4.5 Å
- Optional primary and secondary collimation (Gd-O Soller collimators)
- Oscillating collimator system between sample and detector: 73 Cd coated steel plates of 0.02 cm thickness, angular distance 1.2°
- "Banana" type multidetector, linear position-sensitive BF3 counter (LCC 400CP, Thompson-CSF)
- 400 detectors with angular separation of 0.2°, covering simultaneously a scattering angle range of 79.8°, radius 1.5 m positioning of detector on air cushions for intermediate steps or larger scattering angle range
- SICS (SINQ Instrument Control Software) for control of the measurement
- SEA, TECS (Temperature Control Software) for control of sample environment
- FIT, FULLPROF, ... for data analysis
DMC resolution functions
↑ Top